Commands Listed in Alphabetical Order
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL155 3B:STATus:BIT:TF
NOCARE
sets the st
atus word terminal flag bit value to be used in the search to X.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL155 3B:STATus:BIT:TF? might
return 1.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity
When the MIL-STD-1553 bus search condition is set
to
STATus, this command
specifies the status parity bit value to be used in the search. Returned values are
0, 1, or X (don’t care, which is the default).
SEARCH<x> is the search number,
whichisalways1,and
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MI L1553B:STATus:PARity
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MI L1553B:STATus:PARity?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
2-692 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals