Commands Listed in Alphabetical Order
ZERO sets the va
lue to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL155 3B:STATus:PARity
ONE
sets the parity bit value to be used in the search to
1.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B :STATus:PARity?
might return X, indicating that the parity do
esn’t matter.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit
When the MIL-STD-1553 bus search condition is set to TIMe, this command
specifies either the minimum remote terminal response time (RT) limit
for the amount of time the terminal has to transmit, or it specifies the
minimum inter-message gap (IMG). (You can specify RT or IMG using the
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition TIMe
command.)
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MI L1553B:TIMe:LESSLimit
<NR3>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MI L1553B:TIMe:LESSLimit?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit
Arguments
<NR3> is a floating point number that specifies either the minimum remote
terminal response time (RT) or the inter-message gap (IMG) in seconds.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL155 3B:TIMe:LESSLimit
3.0000e-6
would set either the remote terminal response time (RT) or the
inter-message gap (IMG) to 3.00 μS.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-693