Commands Listed in Alphabetical Order
Following is an
example command sequence showing how to generate a single
search mark on the RF_PHASe waveform when the waveform is logic 0 (less
than 45 degrees):
SEARCH:SEARCH1:TRIGger:A:LOWerthreshol d:RF_PHASe 4.500E+1
SEARCH:SEA
RCH1:TRIGger:A:LOGIc:INPut:RF_PHASe LOW
SEARCH:SEARCH1:STATE 1
SEARCH:SEARCH1:LIST? :SEARC H:SEARCH1:LIST
RF_PHASe,2.4702,2.4702,2.4702,-19.0119 E-6,0.0E+0,0.0E+0,
0.0E+0
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:HIGHLimit
This command specifies the upper limit, in seconds, when searching the record
for pulses whose widths are within or outside of a specified range of two values.
(Use SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:LOWLimit to specify
the lower limit of the range.)
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth: HIGHLimit <NR3>
SEARCH:SEARCH<x>:TRIGge
r:A:PULSEWidth:HIGHLimit?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:WHEn, SEARCH:
SEARCH<x>:TRIGger:A:PULSEWidth:LOWLimit
Arguments
<NR3> is a floating point number that represents the higher value of a range.
Examples
To find all pulses in the waveform record with durations (widths) that fall outside
of the range of 100 nanoseconds to 110 nanoseconds:
SEARCH:SEARCH1:TRIGger:A:PULSEWidth:LO WLimit 100.0E-9
SEARCH:SEARCH1:TRIGger:A:PULSEWidth:HI GHLimit 110.0E-9
SEARCH:SEARCH1:TRIGger:A:PULSEWidth:WH En OUTside
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:LOWLimit
This command specifies the lower limit, in seconds, when searching the record
for pulses whose widths are within or outside of a specified range of two values.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-735