Commands Listed in Alphabetical Order
condition is se
tto
STATus, and the qualifier is set to INrange or OUTrange,
this command specifies the lower limit of the range. (Use the command
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:HIVALue to specify the
upper limit of the range.) The default is all X’s (don’t care).
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADD Ress:VALue <QString>
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADD Ress:VALue?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:QUALifier
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:HIVALue
Arguments
QString is a quoted string of up to 5 characters, where the allowable characters
are 0, 1, and X. The bits specified in the quoted string replace the least significant
bits, leaving any unspecified upper bits unchanged.
Examples
TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRe ss:VALue "01000" sets
the value of the 5–bit remote terminal address to be used in the trigger, when
the qualifier has been set to
LESSthan, MOREthan, E Qual, UNEQual,
LESSEQual
or MOREEQual .
TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRe ss:VALue "01000" sets the
lower limit of the range to 01000, if the qualifier has been set to
INrange or
OUTrange.
TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRe ss:VALue?
might return
“01000”.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:BCR
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word broadcast command received (BCR) bit value (bit 15)
to be used in the trigger. Returned values are 0, 1, or X (don’t care, which is
the default).
B<x>
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-843