Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADD Ress:QUALifier
{LESSthan|MOREthan|EQual|UNEQual|LESSE Qual|MOREEQual|
INrange|OUTrange}
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADD Ress:QUALifier?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:QUALifier
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:HIVALue
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:VALue
Arguments
LESSthan sets the Status Address qualifier to less than.
MOREthan sets the Status Address qualifier to greater than.
EQual sets the Status Address qualifier to equal.
UNEQual sets the Status Address qualifier to not equal.
LESSEQual sets the Status Address qualifier to less than or equal.
MOREEQual sets the Status Address qualifier to greater than or equal.
INrange sets the Status Address qualifier to in range.
OUTrange sets the Status Address qualifier to out of range.
Examples
TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRe ss:QUALifier MOREthan
sets the qualifier to be used with the address field to greater than.
TRIGger:A:BUS:B1:MIL1553B:STATus:ADDRe ss:QUALifier? might
return MORETHAN.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:ADDRess:VALue
When the MIL-STD-1553 bus trigger condition is set to STATus,andthe
qualifier is set to
LESSthan, MOREthan, EQual, UNEQua l, LESSEQual
or MOREEQual, this command specifies the value of the 5–bit remote terminal
address to be used in the trigger. When the MIL-STD-1553 bus trigger
2-842 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual