Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:USB:DATa:HIVALue
This command specifies the binary data string for the upper limit for inside-of-range
and outside-of-range qualifiers for the USB trigger when the trigger condition is
DATAPacket.
Use the command TRIGger:A:BUS:B<x>:USB:DATa:VALue
to set the lower limit.
B<x>
Conditions
This command requires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:USB:DATa:HIVALue <QString>
TRIGger:A:BUS:B<x>:USB:DATa:HIVALue?
Related Commands
TRIGger:A:BUS:B<x>:USB:CONDition,
TRIGger:A:BUS:B<x>:USB:DATa:VALue
Arguments
<QString> within the range 00000000 to 11111111 (00 hex to FF hex).
Examples
TRIGGER:A:BUS:B1:USB:DATA:HIVALUE 00001000” sets the upper limit
to binary 00001000 (08 hex).
TRIGGER:A:BUS:B1:USB:DATA:HIVALUE? might return
:TRIGGER:A:BUS:B1:USB:DATA:HIVALUE “01111111”,which
indicates that the upper limit is 01111111 (7F hex).
TRIGger:A:BUS:B<x>:USB:DATa:OFFSet
This command specifies the data offset in bytes to trigger on. The mi
nimum and
default values are 0 and the maximum is 1024.
B<x>
Conditions
This command requires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Trigger
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-861