Commands Listed in Alphabetical Order
TRIGger:B:BY
Selects or returns whether the B trigger occurs after a specified number of events
or a specified period of time after the A trigger.
Group
Trigger
Syntax
TRIGger:B:BY {EVENTS|TIMe}
TRIGger:B:BY?
Related Commands
TRIGger:B:EVENTS:COUNt, TRIGger:B:TIMe,,
Arguments
EVENTS sets the B trigger to take place following a set number of trigger
events after the A trigger occurs. The number of events is specified by
TRIGger:B:
EVENTS:COUNt.
TIMe sets the B trigger to occur a set time after the A trigger event. The time
period is specified by TRIGger:B:
TIMe.
Examples
TRIGGER:B:BY TIME sets the B trigger to occur at a set time after the A trigger
event.
TRIGGER:B:BY? might return :TRIGGER:B:BY EVENTS indicating that the B
trigger takes place following a set number of trigger events after the A trigger
occurs.
TRIGger:B:EDGE? (Query Only)
Returns the source, slope, and coupling for B trigger.
Group
Trigger
Syntax
TRIGger:B:EDGE?
Related Commands
TRIGger:B:EDGE:COUPling,
TRIGger:B:EDGE:SLOpe,
TRIGger:B:EDGE:SOUrce
Examples
TRIGGER:B:EDGE? might return :TRIGGER:B:EDGE:SOURCE CH1; SLOPE
RISE;COUPLING DC
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-921