Performance Verification
SPG600 & SPG300 Sync Pulse Generators Service Manual 4-5
d. Press the ENTER button.
e. Press the up (
S) or down (T) arrow button to select VIDEO
STANDARD.
f. Press the ENTER button.
g. Press the left (
W) or right (X) arrow button to select 3.58 MHz.
h. Press the ENTER button.
4. Set the frequency counter to trigger on the input.
5. Verify that the displayed frequency is within the range of 3.5795444 MHz to
3.5795464 MHz.
Genlock Function
This test verifies that the genlock function is operating correctly. The following
equipment is required for the test:
TV signal generator (TG700 mainframe with AGL7 module)
Waveform vector monitor
Three 75 Ω BNC cables
Two 75 Ω coaxial terminators
Perform the following procedure to verify that the genlock function is functioning
correctly.
1. Use a 75
Ω BNC cable to connect the REF connector on the sync pulse
generator to the BLACK 1 connector on the TV signal Generator (AGL7
module). See Figure 4-2.
2. Use a 75
Ω BNC cable to connect the other REF connector on the sync pulse
generator to the A connector on the waveform vector monitor. See Figure 4-2.
3. Use a 75
Ω BNC cable to connect the CH 1 connector on the sync pulse
generator to the B connector on the waveform vector monitor. See Figure 4-2.
4. Use the 75
Ω coaxial terminators to terminate the A and B loop through
connectors on the waveform vector monitor rear panel. See Figure 4-2.