Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2-65
CH<x>:DESKew
Sets or queries the horizontal deskew time for this channel. This is equivalent to
setting Deskew in the VERTICAL > Probe Setup menu and entering a value
with the general purpose knob. Deskew lets you compensate for time delay
differences between input channels caused by signals coming in on cables of
different length.
Vertical
CH<x>:DESKew <NR3>
CH<x>:DESKew?
<x> specifies the input channel. Valid values are 1, 2, 3, and 4.
<NR3> is the deskew time for this channel. The range is --10.0 ns to +10.0 ns
with a resolution of 20 ps. Out of range values are clipped.
CH1:DESKEW?
might return ”-4.0E-11”, indicating that the channel 1 deskew setting is
--40.0 ps.
CH<x>:ID? (Query Only)
Returns the probe identification including model name for level II probes.
Returns a null string for non level II probes.
Vertical
CH<x>:ID?
<x> specifies the input channel. Valid values are 1, 2, 3, and 4.
CH1:ID?
might return ”P6243”