TRIGger:B:TYPe { EDGE }
TRIGger:B:TYPe?
EDGE is a normal trigger. A B trigger event occurs when a signal passes through a
specified voltage level in a specified direction and is controlled by the TRIG-
ger:B:EDGE commands.
TRIGGER:B:TYPE?
returns :TRIGGER:B:TYPE EDGE indicating that the B trigger type is a edge
trigger.
TRIGger:DELay Commands
The TRIGger:DELay commands listed in Table 2--42 have the same function and
syntax as the equivalent TRIGger:B commands, and are for compatibility with
other Tektronix instruments.
Table 2- 42: TRIGger:DELay commands
TRIGger:DELay TRIGger:DELay: BY
TRIGger:DELay: EDGE:COUPling TRIGger:DELay: EDGE:SLOpe
TRIGger:DELay: EDGE:SOUrce TRIGger:DELay:EVENTS:COUNt
TRIGger:DELay: LEVel TRIGger:DELay: SETLevel
TRIGger:DELay:STATE TRIGger:DELay:TIMe
TRIGger:DELay: TYPe
TRIGger:EXTERnal:PRObe
(TDS3014B/C, TDS3024B, TDS3034B/C, TDS3044B, TD S3054B /C, TDS3064B Only)
Sets or returns the gain factor value of the external probe connector.
Trigger
TRIGger:EXTERnal:PRObe <NR1>
TRIGger:EXTERnal:PRObe?
Syntax
Arguments
Examples
Group
Syntax