Command Descriptions
2- 100
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
DIAg:LOOP:OPTion
Sets the diagnostic loop option.
Calibration and Diagnostic
DIAg:LOOP:OPTion {ALWAYS|FAIL|ONFAIL|ONCE}
DIAg:LOOP:OPTion?
ALWAYS continues looping until diagnostics are stopped via the front panel or by
an instrument command.
FAIL causes looping until the first diagnostic failure or until diagnostics are
stopped.
ONFAIL causes looping on a specific test group as long as a FAIL status is
returned from the test.
ONCE executes diagnostics test sequence once.
DIAG:LOOP:OPTION FAIL
stops the diagnostic loop after the first failure.
DIAg:LOOP:STOP (Set Only)
Stops diagnostic at the end of the current loop.
Calibration and Diagnostic
DIAg:LOOP:STOP
DIAG:LOOP:STOP
stops diagnostic at the end of the current loop.
Group
Syntax
Arguments
Examples
Group
Syntax
Examples