Command Descriptions
2- 270
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
LINE sets the trigger source to the AC line frequency.
NOTE. LINE is not available when the instrument is operating on battery power.
ALTernating sets the trigger source to alternate between each displayed input
channel.
TRIGGER:A:EDGE:SOURCE LINE
specifies the AC line voltage as the A edge trigger source.
TRIGGER:A:EDGE:SOURCE?
might return :TRIGGER:A:EDGE:SOURCE CH2 for the A edge trigger source.
TRIGger:A:HOLdoff? (Query Only)
Returns the A trigger holdoff time.
Trigger
TRIGger:A:HOLdoff?
TRIGGER:A:HOLDOFF?
might return :TRIGGER:A:HOLDOFF:TIME 2.508E-7.
TRIGger:A:HOLdoff:TIMe
Sets or queries the A trigger holdoff time. This is equivalent to setting Hold off
(Time) in the Mode & Holdoff side menu.
Trigger
TRIGger:A:HOLdoff:TIMe <NR3>
TRIGger:A:HOLdoff:TIMe?
TRIGGER:A:HOLDOFF:TIME 2.5e-7
to set the holdoff value to 250.8 ns.