Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 309
VERBOSE ON
sets the Verbose State true.
VERBOSE?
might return the value 1, showing that the Verbose State is true.
HEADer, *LRN?, SET?
*WAI (Set Only)
(Wait) Prevents the oscilloscope from executing further commands or queries
until all pending operations finish. This command allows you to synchronize the
operation of the oscilloscope with your application program. Synchronization
methods are described on page 3--7.
Status and Error
*WAI
BUSY?, *OPC
WAVFrm? (Query Only)
Returns WFMPre? and CURVe? data for the waveform specified by the
DATa:SOUrce command. This command is equivalent to sending WFMPre?;
CURVe?. If the waveform specified by the DATa:SOUrce command is not
displayed, the oscilloscope returns only the waveform transmission parameters
(BYT_Nr, BIT_Nr, ENCdg, BN_Fmt, BYT_Or).
The WFMPre response is first, followed by the CURVe response with an
intervening semicolon.
Waveform
WAVFrm?
See WFMPre? and CURVe? commands.
CURVe, DATa:SOUrce, WFMPre?
Examples
Related Commands
Group
Syntax
Related Commands
Group
Syntax
Returns
Related Commands

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals