Specifications (Cont.)
Vertical
N
u
m
b
er
o
f
c
hann
el
s TDS3012, TDS3032, TDS3052 TDS3014, TDS3034, TDS3054
2 plus external trigger input 4
Digitizers 9-bit resolution, separate digitizers for each channel sample
simultaneously
SCA
LE Rang
e
tNC
1 MW 50 W
(a
t
B
NC
)
1 mV/div to 10 V/div
1mV/div to 1 V/div
Fine SCALE Adjustable with 1% resolution
Polarity Normal and Invert
Position range ±5 divisions
A
na
lo
g bandw
i
d
t
h,
50 W
lot icl t W
Bandwidth limit set to Full, operating ambient 30 °C, derate
1%/°C above 30 °C
(a
l
s
o
t
yp
ic
a
l
a
t
1 MW
with standard probe)
Scale range
TDS3012
TDS3014
TDS3032
TDS3034
TDS3052
TDS3054
5 mV/div to
1 V/div
100 MHz 300 MHz 500 MHz
2 mV/div to
4.98 mV/div
100 MHz 250 MHz 300 MHz
1 mV/div to
1.99 mV/div
90 MHz 150 MHz 175 MHz
Calculated rise time,
typical
3.5 ns 1.2 ns 0.7 ns
Analog bandwidth limit,
typical
Selectable between 20 MHz, 150 MHz (not available on
TDS3012 or TDS3014), or Full
Lower frequency limit,
AC coupled, typical
7 Hz for 1 MW, reduced by a factor of ten when using a 10X
passive probe; 140 kHz for 50 W

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals