Appendix E: Performance Verification
E–12
TDS3000 Series User Manual
24. Verify that the peak-to-peak measurement is 425 mV.
25. Push the waveform off button.
26. Repeat steps 11 through 25 for each channel of the oscilloscope
(not including the external trigger input on some models).