EasyManuals Logo
Home>Tektronix>Test Equipment>TDS5000B Series

Tektronix TDS5000B Series Specification And Performance Verification Technical Reference

Tektronix TDS5000B Series
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
Performance Tests
2-58
TDS5000B Series Specifications and Performance Verification
H CHECK that the readout C H 2 Pk-Pk is between 100 mV and
150 mV, inclusive.
H Enter the voltage in the test record.
4. Disconnect the hookup: Disconnect the cables from the channel inputs and
the rear panel outputs.
Equipment
required
One BNC male-to-male cable and a BNC female-to-clip lead adapter
used between CH1 and the Probe Comp Output
DC Calibrator
Prerequisites See page 2--17. Also, the oscilloscope must have passed Check
Long-Term Sample Rate and Delay Time Accuracy on page 2--39.
1. Install the test hookup and preset the instrument controls:
a. Hook up test-signal: Connect the PROBE COMP output to CH 1 using
the BNC male-to-male cable and the BNC female-to-clip lead adapter.
See Figure 2--24.
TDS5000B Series oscilloscope
BNC cable and adapter
Figure 2- 24: Initial test hookup
b. Initialize the oscilloscope: Press the DEFAULT SETUP button.
c. Modify the initialized front-panel control settings:
H Set the Vertical SCALE to 200 mV.
H Set the Horizontal SCALE to 200 s.
H Press PUSH TO SET 50%.
H Use the Vertical POSITION knob to center the display on the
screen.
H From the toolbar bar, click Horiz and select the Acquisition tab.
Check Probe
Compensation Output

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals