Removal and Installation Procedures
TLA5000 Series Service Manual
6-35
Daughter Boards
1. Remove the trim and covers by following the procedure on page 6--7.
2. Remove the bottom covers.
3. Set the instrument so its top is down on the work surface and its bottom is
facing up.
4. Locate the daughter board to be replaced.
5. Remove the three T-15 Torx-drive screws securing the daughter boards to the
Acquisition board.
6. Lift the daughter board off the connector.
7. To reinstall the daughter board place the daughter board over the connector
and gently press it in place.
8. Complete steps 1 through 5 in reverse order. Tighten the T-15 Torx-drive
screws to 8-in lbs.
Verifying Operation
To verify the proper operation of the logic analyzer, follow these steps:
1. Plug the power cord in and turn the logic analyzer on.
2. Verify that the instrument passes all power-up diagnostics.
3. Exit all applications and close any open windows.
4. Click Start in the Windows tool bar.
5. Select Programs from the Start menu.
6. Select the CheckIt Utilities application from the Programs menu.
7. Run the appropriate verification tests from the application.
8. Turn the logic analyzer off, and then on again before running any application
software including the TLA application software.