AFA4000/E VAV Airflow Controller / p.122
Menu item
Option
Instructions
Sash Warning
Timer (cont.)
Sash warn rpt time
If the alarm is enabled, set the repeat time.
Factory Settings
Enter password
Use / to enter password, then press
Mute.
Restore factory setting
Press Mute to restore the factory settings.
Press to return to the Monitor config
menu (section 8.1.1).
Done
Return to Monitor set up
menu.
8.1.1.1 Cal config menu
Note: Use / to increase or decrease parameter values, then press Mute to save the new
value.
Menu item
Instructions
Display Units
Select the required display units:
m/sec
fpm
Low Air Alarm
Set the low air flow alarm threshold.
Low Air Cutoff
Enable or disable the low air cut-off.
If enabled, set the low air flow threshold.
High Air Cutoff
Enable or disable the high air cut-off.
If enabled, set the high air flow threshold.
Warning Air
Set the air flow warning threshold.
Warning Air Reset
Set the air flow warning reset level.
High Air Alarm
Enable or disable the high air alarm.
If enabled, set the high air flow threshold.
Lower Air Fluctuation
Set the lower air fluctuation level (%).
Higher Air Fluctuation
Set the higher air fluctuation level (%).
Low/High Air Difference
Set the low / high air difference value.