CC-10
INSTRUMENT MANUAL
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
CC-10 MANUAL REV D
Page 30 of 40
Y
TELEVA
1. gauge to host
1 2 3 4 5 6 7 8
< STX>
Address
(HEX)
N
Error code
< CR>
2.
error response code
Error code
Explanation
0001
Command error undefined character (other than R, W, C or S) in
CMD line
0002
Mode error undefined numbers in MODE line
0003
Data error undefined numbers in DATA line
0004
Gauge is busy for it is in parameter set mode (when W or C
command is sent)
0005
Gauge is uncontrollable (only S command mode 7 is acceptable)
S (Status) command: Read pressure data (mode 1)
Unit address = 0
Host to gauge
< STX>
Address
(HEX)
0 to F
S
Mode
< CR>
Example code:
0X02
0X30
0X53
0X31
0X0D
-
0
S‟
1
-
Error response
CODE EXAMPLE
Hex
ASCii