CC-10
INSTRUMENT MANUAL
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
CC-10 MANUAL REV D
Page 38 of 40
10.0
E
rr
o
r
m
e
ss
a
g
e
li
s
t
Atmosphere adjustment is tried at
inadequate pressure (perhaps in
vacuum)
Make sure the gauge is at
atmosphere before adjustment
Zero adjustment is tried at
insufficient vacuum
Perform the zero adjustment in
sufficient vacuum
Crystal sensor oscillation error
Needs crystal sensor replacement.
Contact service agent.
Needs electronics repair. Contact
service agent
A/D converter calibration error
Needs electronics repair. Contact
service agent
`
EE Error
(scroll display)`
Needs electronics repair. Contact
service agent
11.0
T
r
oub
l
e
s
hoo
t
i
ng
Check connector. Check power (24
VDC) supply
Pressure reading is
unstable
Contamination in the
sensor
Pressure reading and HV
LED are blinking
Cold cathode sensor
is not ignited
(possibly the
pressure is low
when power is
turned on)
In low pressures cold cathode
sensor tend to take a little long time
to ignite. Wait until the pressure
reading appears.
Pressure display does not
decrease under 7.5E-3 Torr
Perform manual zero adjustment
when the pressure is below 4E-5
Torr