EasyManua.ls Logo

TESTO 320 - Measurement view; Flue gas

TESTO 320
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Using the product
45
6.2.1.4. Measurement view
Only those measurement parameters and -units that are enabled in
the reading display appear in the reading display, in the saved
measurement records and on record printouts.
> Before carrying out measurements, set up the measured value
display in such a way that the required parameters and units
are enabled, see Measurement view, page 34.
6.2.1.5. Setting the location and fuel
Before carrying out measurements, the location and fuel must be
correctly selected, see Address/Location, page 30 and see Fuels,
page 42.
6.2.2. Flue gas
To achieve usable measurement results, the measurement
period of a flue gas measurement should be approx. 3 min
and the measuring instrument should display stable
measuring values.
Call up function:
1. [ ] Measurement options [OK] Flue Gas [OK].
2. Select the fuel → [OK].
Carrying out the measurement:
1. Start measurement: [ ].
If a separate measurement of CO undiluted has not yet
been carried out, this value is calculated using the
measured values of the flue gas probe and is updated
continuously.
If CO undiluted and/or a draught measurement has already
been carried out separately, the value obtained is applied.
- The measured values are displayed.
2. End measurement: [ ].
Options
> [Options] Clipboard: data is saved to the clipboard.
> [Options] Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] Save: the measured values are saved in a record.
> [Options] Fluegas matrix: the measured values are
displayed as a flue gas matrix, see below.
1.888.610.7664 info@Testo-Direct.com
www.Testo-Direct.com

Table of Contents

Related product manuals