6 Using the product
51
Determine smoke tester no./smoke nos./oil depos. with the
smoke tester testo 308 and transmit wirelessly:
- The testo 308 must be in data transfer mode ( lights up).
> [Options] t308.
- The values recorded by the smoke tester are transferred to the
testo 320.
Entering the heat carrier temperature:
> Heat carrier. [Edit] → enter value → [OK].
Options:
> [Options] Clipboard: data is saved to the clipboard.
> [Options] Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] Save: the measured values are saved in a record.
> [Options] Reset values: The entered values are deleted.
6.2.9. Pressure
The gas pressure set (0554 1203) must be connected.
Call up function:
> [ ] Measurement options [OK] Pressure [OK].
Carrying out the measurement:
Dangerous mixture of gases
Danger of explosion!
> Make sure there are no leaks between the sampling point and
the measuring instrument.
> Do not smoke or use naked flames during measurement.
Do not measure for longer than 5 min, as the drift of the
pressure sensor could mean that the measured values are
outside the tolerance limits.
1. Connect the gas pressure set to the probe socket.
2. Start measurement: [ ].
- Pressure zeroing is carried out (system must be unpressurised).
3. Pressurise the system.
- The measured value is displayed
4. End measurement: [ ].