7 Maintaining the product
61
5. Install new sensor / new bridge in the slot.
6. Push the hose connections on to the sensor/bridge.
7. Refit and close the service cover.
After replacing an O2 sensor, wait for an equalisation
period of 60 min to elapse before using the instrument
again.
When retrofitting a sensor, the associated measurement
parameter and unit must be enabled, see Sensor settings,
page 40.
7.5. Recalibrating/adjusting sensors
See Recalibration/adjustment, page 41.
7.6. Modular flue gas probe
7.6.1. Cleaning the flue gas ducts
✓ Disconnect the flue gas probe from the measuring instrument.
2. Blow compressed air through the flue gas ducts in probe
module and probe handle (see illustration). Do not use a brush!
3. Fit a new probe module on the handle and engage in place.
7.6.2. Replacing the probe module
✓ Disconnect the flue gas probe from the measuring instrument.