Instrument settings and analysis
28
Selecting the indicator type and bottle size
Specifying the measuring/monitoring range of Testomat
®
808
The measuring/monitoring range of Testomat
®
808 is specified by
the selected indicator type only.
Please refer to the section entitled “Testomat
®
808 spare parts and
accessories” on page 41 for a list of available indicators.
All listed indicator types are available in two bottle sizes. To specify
the size of the indicator bottle, move slide switch T2 on the rear of the
controller board to the respective switch position (see the section
entitled “Internal design Testomat
®
808” on page 16):
Switch position
The status of slide switch T2 is read after resetting the indicator
display to 100% and after a reset.
Set indicator quantity to 100%
After selecting the bottle size and inserting the bottle, press key 4
(horn) to set the display for the indicator quantity to 100%.
Carrying out an analysis
After switching it on, the instrument commences automatic interval
mode. The first analysis starts after 15 seconds. The following anal-
yses start automatically according to the set interval pause.
Duration of the interval pause
The duration of the interval pause between two measurements can
be set between 0 and 60 minutes, also see the section entitled
“Interval pause” on page 31.
Attention! After a poor analysis, the interval pause is ignored in
some switching positions of function key S6 (2/3/5/6/8 and 9) and
another analysis is carried out immediately.