Further basic functions and settings
31
External flush process:
Long press (approx. 2 seconds) key 2 (Flush) to activate the exter-
nal flush valve.
The external valve is activated via relay K1 and the line is flushed.
Long key press (approx. 2 seconds) key 2 (Flush) again to termi-
nate the external flush process.
The LED above key 2 (Flush) and LED K1 illuminate as long as the
external flush valve is activated.
Interval pause
If the analysis is started via a timer, the interval between two anal-
yses (plus flush time) is determined by the interval pause. The short-
est interval can be 0 minutes. In this case, analyses are carried out
continuously. The longest interval is 60 minutes.
Set the desired measuring interval pause via function key S6 on the
controller board (see the section entitled “Rear of the controller
board” on page 16). The following switching positions and interval
pauses are possible:
Position
Interval pause
0
0 min.
1
5 min.
2
10 min.
3*
15 min.
4
20 min.
5
25 min.
6
30 min.
7
40 min.
8
50 min.
9
60 min.
* Delivery status
Switch position
The current switch position is read after evaluating a measuring
result and after a reset.
Resetting the interval pause time
However, if the standby function has been activated and subse-
quently deactivated during an analysis pause, the interval pause
time is reset.
NOTE
NOTE