6 | Operating Standards Ultra Low Temperature Freezers
Operating Standards
The freezers described in this manual are classified for use as
stationary equipment in a Pollution Degree 2 and over voltage
Category II environment.
These units are designed to operate under the following
environmental conditions:
• Indoor use
• Altitude up to 2000 m
• Maximum relative humidity 60% for temperatures within
15°C to 32°C (59°F to 90°F). For TSX70086* models,
the maximum operating environment is 28°C (83°F).
• Main supply voltage fluctuations not to exceed
±10% of the nominal voltage.
• For the TSX series, the ULT should not be connected to a
GFCI (Ground Fault Circuit Interrupter) protected outlet as
it may be subject to nuisance tripping.
Electrical
The last character in the model number listed on the dataplate
identifies the electrical specifications for your unit. Specific
unit current rating is listed on the dataplate.
The voltage types are A, D, V and G as specified in the
following table:
Table 1. TSX Series Electrical Specifications
Model Voltage Frequency Current
400D/V 208-230 V 50/60 Hz 4.0 A
400A 115 V 60 Hz 9.0 A
400G 100 V 50/60 Hz 10.5 A
500D 208-230 V 60 Hz 5.1 A
500V 208-230 V 50 Hz 5.2 A
500A 115 V 60 Hz 9.5 A
500G 100 V 50/60 Hz 9.5 A
600D/V 208-230 V 50/60 Hz 4.0 A
600A 115 V 60 Hz 8.5 A
600G 100 V 50/60 Hz 9.5 A
700D 208-230 V 60 Hz 5.8 A
700V 208-230 V 50 Hz 6.1 A
700A 115 V 60 Hz 10.6 A