Ultra Low Temperature Freezers Installation | 7
Installation
Location
Install the unit in a level area free from vibration with a
minimum of 8 inches (20 cm) of space on the top and sides,
6 inches (15 cm) in back. Refer Leveling for further
instructions on leveling cabinets. Allow enough clearance so
that door can swing open at least 85°.
The rear spacing posts provided with the freezer can be used
to ensure proper clearance. To install the spacing posts,
screw them into the back in the rear deck area.
Do not position the equipment in direct sunlight or near
heating diffusers, radiators, or other sources of heat. The
ambient temperature range at the location must be 15°C to
32°C (59°F to 90°F). For TSX70086* models, the maximum
operating environment is 28°C (83°F).
Wiring
Leveling
Ensure that the floor is level. The unit must be level both front
to back and side to side.
TSX40086* models are equipped with one leveling leg on the
right hand side. These may be used to help prevent the unit
from shifting during a door opening.
Be certain to lock the brakes for units equipped with casters.
WARNING: Do not exceed the electrical rating
printed on the data plate located on the lower
left side of the unit.
CAUTION: Connect the equipment to the
correct power source. Incorrect voltage can
result in severe damage to the equipment.
CAUTION: For personal safety and
trouble-free operation, this unit must be
properly grounded before it is used. Failure to
ground the equipment may cause personal
injury or damage to the equipment. Always
conform to the National Electrical Code and
local codes. Do not connect the unit to
overloaded power lines.
CAUTION: Do not position the unit in a way
that impedes access to the disconnecting
device or circuit breaker in the back of the unit.
CAUTION: Always connect the freezer to a
dedicated (separate) circuit. Each freezer is
equipped with a service cord and plug
designed to connect it to a power outlet which
delivers the correct voltage. Supply voltage
must be within ±10% of the freezer rated
voltage.
CAUTION: Never remove or disable the
grounding prong from the service cord plug.
If the prong is removed, the warranty is
invalidated.

Other manuals for Thermo Scientific TSX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSX Series and is the answer not in the manual?

Thermo Scientific TSX Series Specifications

General IconGeneral
TypeLaboratory Refrigerator
Cooling TypeForced air cooling
Defrost TypeAutomatic
Door TypeSolid or glass door options
Controller TypeMicroprocessor
Certifications/ComplianceUL, CE
Energy ConsumptionVaries by model; refer to product specification sheet
Voltage115V
Exterior Dimensions (H x W x D)Varies by model; refer to product specification sheet
Interior Dimensions (H x W x D)Varies by model; refer to product specification sheet
WeightVaries by model; refer to product specification sheet
ShelvesAdjustable shelves
Energy EfficiencyEnergy Star certified models available