46 | Appendix A: Alarm Summary Ultra Low Temperature Freezers
Appendix A: Alarm Summary
Alarm Summary
Alarm Message Tone
Heart
Status
Remote
Alarm
Event
Ringback
*
Message Details
Warm Alarm Active High Red Yes Yes
Message Header: Warm Alarm
Message Details: The freezer temperature
has exceeded the warm alarm set point of
XX.X C.
Prolonged door openings and warm product
loading may cause warm alarms.
Ticker Message: Warm Alarm active.
Press bell icon for more information.
Warm Alarm
Inactive
None Yellow None None
Message Header: Warm Alarm Recovered
Message Details: The freezer temperature
has recovered from a warm alarm event.
Select check box and press Acknowledge to
clear this notification.
Ticker Message: None
Cold Alarm Active High Red Yes Yes
Message Header: Cold Alarm
Message Details: The freezer temperature
has exceeded the cold alarm set point of
XX.X C
Ticker Message: Cold Alarm active.
Press bell icon for more information.
Cold Alarm Inactive None Yellow None None
Message Header: Cold Alarm Recovered
Message Details: The freezer temperature
has recovered from a cold alarm event.
Select check box and press Acknowledge to
clear this notification.
Ticker Message: None
Door Open Ajar High Red Yes Yes
Message Header: Door Open Alarm
Message Details: Door open for greater than
3 minutes will cause door open alarm.
Ticker Message: Door Open Alarm active.
Press bell icon for more information.