Page #15 background image
10 | Start Up Ultra Low Temperature Freezers
Start Up
Connectivity
Requirements
There are two requirements to take full advantage of your
freezers connectivity options:
1. The freezer will need to be connected to a wireless
network with internet connection, the freezer is only able
to use a wireless connection.
a. Refer Table 2 for acceptable network security param-
eters.
b. Speak with your local Information Technology (IT)
group about the correct wireless network and
password to use.
2. The individual monitoring the unit will need a
Thermo Fisher Connect account and will need to link the
unit to his/her account via InstrumentConnect™.
To create a Thermo Fisher Connect account:
a. Use a web browser to open the following URL:
https://apps.thermofisher.com
b. Choose "Create an Account" and follow the
instructions to establish a new account.
(Remember your login information for later use.)
c. (Optional) Download the InstrumentConnect
mobile/tablet application from the AppStore or
PlayStore.
Note: InstrumentConnect™ is the section of Thermo Fisher
Connect where all connected instruments can be monitored.
To access this section, click the following InstrumentConnect
icon indicated.
Note: If you are a user in China, the web client allows you to
switch to that region after logging in or you can log in directly
to the China environment by using
https://china.apps.thermofisher.com.
Specifications
Note: Thermo Fisher Scientific connected devices support
certification validation.
In case of Failure
If your freezer is unable to connect to either a wireless
network or a cloud account, verify your name and password
and attempt to reconnect.
If issues persist, contact your Local Support.
Note: If your network password expires or changes, your
freezer will no longer connect automatically. Ensure to
manually update your password in the freezer user interface
settings to reconnect. Failure to reconnect within 3 days will
result in data loss.
Table 2. Specifications
Parameter Specification
Default Wi-Fi
transmission frequency
to the cloud.
Every 5 minutes for sensor
data transmission.
Real-time for alarm events.
Wi-Fi protocol
2.4 GHZ only: IEEE 802.11B,
IEEE 802.11G, IEEE 802.11N
Wi-Fi Security
WPA2PSK
WPA2PEAP(PEAP)
Wi-Fi Data Rate Minimum of 1 Mbps
Wi-Fi Range Up to 30 meters
Minimum Required
Wireless Signal
-67 dbM
Firewall Ports that must
open
123, 443

Other manuals for Thermo Scientific TSX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSX Series and is the answer not in the manual?

Thermo Scientific TSX Series Specifications

General IconGeneral
TypeLaboratory Refrigerator
Cooling TypeForced air cooling
Defrost TypeAutomatic
Door TypeSolid or glass door options
Controller TypeMicroprocessor
Certifications/ComplianceUL, CE
Energy ConsumptionVaries by model; refer to product specification sheet
Voltage115V
Exterior Dimensions (H x W x D)Varies by model; refer to product specification sheet
Interior Dimensions (H x W x D)Varies by model; refer to product specification sheet
WeightVaries by model; refer to product specification sheet
ShelvesAdjustable shelves
Energy EfficiencyEnergy Star certified models available

Related product manuals