2.15 Setup and function of temperature controllers
ITC500 / page 46
2.15.4.3 D-share
4. Change the set temperature repeatedly between values of ambient
temperature +2°C and ambient temperature -2°C and observe the settling
behavior of the actual temperature.
Increase the D-share gradually by turning the corresponding potentiometer
clockwise.
Higher values will decrease the amplitude of the overshoots.
The D-share has been set correctly when the actual temperature remains
stable near the set temperature after a minimum of overshoots.
2.15.4.4 I-share
5. Change the set temperature repeatedly between values of ambient
temperature +2°C and ambient temperature -2°C and observe the settling
behavior of the actual temperature.
Increase the I-share gradually by turning the corresponding potentiometer
clockwise.
Higher values will accelerate the settling to the set temperature.
The I-share has been set correctly when the actual temperature settles in
shortest time to the set temperature.