2.23 Laser voltage measurement
ITC500 / page 55
3 Remote control by a PC (ITC5xx-IEEE)
NOTE
The following operating elements are still working with the remote control
active:
The display
The display selection
all potentiometers except the main dial knob
NOTE
All analog values are read and written in SI units, i.e. A (not mA), W (not
mW) etc. Letters may be small or capital letters.
Attention
Before programming an ITC5xx first set the limit value of the laser
diode current I
LD LIM
(hardware limit) for the applied laser diode with a
screwdriver.
The corresponding potentiometer is marked ADJ and located next to
the LED I
LD LIM
situated at the front of the ITC5xx.
The value I
LD LIM
is constantly measured and can be checked if the
LED I
LD LIM
selects the display of the limit current.