3.5 Status command group
ITC500 / page 70
3.5 Status command group
(Refer to chapter 3.8, "Status reporting" on page 90)
3.5.1 Query Device-Error-Condition-Register (DEC)
Syntax: ":STAT:DEC?"
Response: [:STAT:DEC <NR1>]
Description: Queries the Device-Error-Condition-Register (DEC) and returns the
content in decimal notation. The content is not modified.
3.5.2 Query Device-Error-Event-Register (DEE)
Syntax: ":STAT:DEE?"
Response: [:STAT:DEE <NR1>]
Description: Queries the Device-Error-Event-Register (DEE) and returns the
content in decimal notation. The content is cleared.
3.5.3 Device-Error-Event-Enable-Register (EDE)
Programming:
Syntax: ":STAT:EDE <NR1>"
Valid Range: 0...65535
Default Value: 0
Description: Sets the Device-Error-Event-Enable-Register (EDE).
Reading:
Syntax: ":STAT:EDE?"
Response: [:STAT:EDE <NR1>]
Description: Queries the Device-Error-Event-Enable-Register (EDE) and returns
the content in decimal notation. The content is not modified.