EasyManuals Logo

THORLABS ITC502 User Manual

THORLABS ITC502
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
3.3 Common commands and queries
ITC500 / page 67
3.3.9 Service-Request-Enable-Register (SRE)
Programming:
Syntax: "*SRE <NR1>"
Valid Range: 0..255
Default Value: 0
Description: Sets the Service-Request-Enable-Register (SRE).
Reading:
Syntax: "*SRE?"
Response: [<NR1>]
Description: Queries the Service-Request-Enable-Register (ESE) and returns the
content in decimal notation. The content is not modified.
3.3.10 Query Status-Byte-Register (STB)
Syntax: "*STB?"
Response: [<NR1>]
Description: Queries the Status-Byte-Register (STB) and returns the content in
decimal notation. Bit 6 (MSS) is set to 0, the other bits keep
unchanged.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the THORLABS ITC502 and is the answer not in the manual?

THORLABS ITC502 Specifications

General IconGeneral
BrandTHORLABS
ModelITC502
CategoryController
LanguageEnglish

Related product manuals