3.6 ITC5xx specific commands
ITC500 / page 80
3.6.15 Temperature Sensor (SENS)
Programming:
Syntax: ":SENS AD" (AD590 family)
":SENS THL" (thermistor, 20 kΩ range)
":SENS THH" (thermistor, 200 kΩ range)
Assumption: The TEC output is switched off.
Description: Selects the type of temperature sensor. Sets the programmed
temperature / resistance to default values.
Reading:
Syntax: ":SENS?"
Response: [:SENS AD]
[:SENS THL]
[:SENS THH]
Description: Queries the type of the temperature sensor.
3.6.16 Switching the TEC on and off (TEC)
Programming:
Syntax: ":TEC OFF"
":TEC ON"
Default Value: OFF
Assumption: To switch the TEC on there must be no device errors (no sensor, TEC
open, over temperature, ...).
Description: Switches the TEC output on or off.
Reading:
Syntax: ":TEC?"
Response: [:TEC OFF]
[:TEC ON]
Description: Queries the state of the TEC output.