1-3
■
- Do not hold the head, chin rest, forehead pad, joystick, or cables when
moving the instrument. These components are detachable and the
instrument may drop, resulting in injuries.
- Install the instrument in a location not subject to direct sunlight, high
temperature and humidity, or air containing dust, salt, and/or sulfur.
Otherwise, failure or malfunction may occur.
- Install the instrument in a stable, level location free of vibration or
mechanical impact. Otherwise, measurement cannot be conducted
correctly. Also, the instrument may topple over or fall down, resulting in
fire or serious accident.
- Place the instrument between the patient and physician so that they can
face each other.
- Install the instrument in a location with ample clearance from other
devices to allow smooth inspection.
- Check the frequency, voltage, and allowable current (or power
consumption) of the power source.
- The AC adapter for the touch panel LCD is different from that for the
external HDD. Connect the correct adapter. Incorrect connection may
result in failure of the instrument.
■
Precautions before operation
- Check that the instrument works properly by inspecting switch and
button operations, display, and printing.
- Check that all cables are connected correctly.
- Since simultaneous use of multiple devices can cause misdiagnosis or
result in a hazardous situation, exercise caution when using this
instrument.
- Check the sections that the patient will directly touch.
- Peel off the top sheet of chin rest paper and clean the forehead pad with
a cloth dampened with alcohol before measurement.
- Check that the instrument is correctly grounded.