26
down to 10% of the current range.
Step 4: Set the integration time.
The setting of the integration time affects the measurement speed and measurement accuracy. The longer the
integration time, the higher the accuracy but the slower the measurement. 1PLC, 10PLC, 100PLC can inhibit the
power supply noise. Selecting 100PLC can provide the best noise rejection, but at a slower rate.
4.2.8 Capacitance Measurement
Step 1: Configure the test leads as follows:
Step 2: Press the button on the front panel to enter the capacitance test interface.
Step 3: Select the range.
Press Range and select the range. Auto range (Auto) can automatically selects the appropriate range for