55
READ?
Typical response: +4.27230000E+00, +4.27150000E+00, +4.27190000E+00, +4.27170000E+00,
+4.27200000E+00
8.2.3.9 TRIGger:DELay:AUTO
Description: Disable or enable automatic trigger delay. If enabled, the instrument sets the delay based on function,
range, and integration time or bandwidth.
By default, TRIGger:DELay:AUTO is ON.
Syntax: TRIGger:DELay:AUTO {ON|1|OFF|0}
TRIGger: DELay: AUTO?
Parameters: {ON|1|OFF|0}
Example: Returns the result of five DC voltage measurements. There is an automatic delay between each
measurement
CONF:VOLT:DC 10
SAMP: COUN 5
READ?
Typical response: +4.27230000E+00, +4.27150000E+00, +4.27190000E+00, +4.27170000E+00,
+4.27200000E+00
8.2.3.10 SAMPle:COUNt
Description: Specifies the number of measurements (number of samples) taken each time the instrument is
triggered.
You can use the specified sample count with a trigger count (TRIGger:COUNt). The total number of
measurements returned before returning to the "idle" trigger state is the product of the sample count and the
trigger count.
Syntax: SAMPle:COUNt {<count>|MIN|MAX|DEF}
SAMPle:COUNt? [{MIN|MAX|DEF}]
Parameters: Number of samples per trigger
Example: Return ten sets of four-time resistance measurements and start each measurement group with an
external trigger:
CONF:RES 1E6
SAMP: COUN 4
TRIG: COUN 10
TRIG: SOUR EXT
READ?
Typical response: +1.00520000E+06, ... (40 measurements)
8.2.4 [SENSe:]VOLTage subsystem
This subsystem configures AC voltage measurement, DC voltage measurement.