49
8.2.1.9 CONFigure:FRESistance
Description: Configure the instrument for four-wire resistance measurement, setting all measurement parameters
and trigger parameters to default values
Syntax: CONFigure:FRESistance [{<range>|AUTO|MIN|MAX|DEF}]
Parameters: <range>: {10|100|1k|10k|100k|1MA|10MA|100MA} Default: AUTO
Example: CONF: FRES 10MA
8.2.1.10 CONFigure:FREQuency
Description: Configure the instrument for frequency measurement, set all measurement parameters and trigger
parameters to default values
Syntax: CONFigure:FREQuency [{<range>|MIN|MAX|DEF}]
Parameters: <range>: {100m|1|10|100|750} Default: AUTO
Example: CONF:FREQ 10
8.2.1.11 CONFigure:PERiod
Description: Configure the instrument for periodic measurements, setting all measurement parameters and trigger
parameters to default values
Syntax: CONFigure:PERiod [{<range>|MIN|MAX|DEF}]
Parameters: <range>: {100m|1|10|100|750} Default: AUTO
Example: CONF:PER 10MA
8.2.1.12 CONFigure:TEMPerature
Description: Configure the instrument for temperature measurement and set all measurement parameters and
trigger parameters to default values
Syntax: CONFigure:TEMPerature [{FRTD|RTD|FTHermistor|THERmistor}]
Parameters: <probe_type>: FRTD|RTD|FTHermistor|THERmistor
Example: CONF:TEMP RTD
8.2.1.13 CONFigure:CAPacitance
Description: Configure the instrument for capacitance measurement, setting all measurement parameters and
trigger parameters to default values
Syntax: CONFigure:CAPacitance [{<range>|AUTO|MIN|MAX|DEF}]
Parameters: <range>:{1n|10n|100n|1u||10u|100u|1m|10m} Default: AUTO
Example: CONF:CAP 10u
8.2.2 MEASure subsystem
With MEASure? queries, you can use a single command to select functions and ranges. All other parameters are