48
Example: CONF:VOLT:AC 100
8.2.1.4 CONFigure:CURRent:DC
Description: Configure the instrument for DC current measurement, setting all measurement parameters and
trigger parameters to default values
Syntax: CONFigure:CURRent:DC [{<range>|AUTO|MIN|MAX|DEF}]
Parameters: <range>: {100 μA|1 mA|10 mA|100 mA|1 A|3 A|10 A}. Default: AUTO
Example: CONF:CURR:DC 100m
8.2.1.5 CONFigure:CURRent:AC
Description: Configure the instrument for AC current measurement, setting all measurement parameters and
trigger parameters to default values
Syntax: CONFigure:CURRent:AC [{<range>|AUTO|MIN|MAX|DEF}]
Parameters: <range>: {100 μA|1 mA|10 mA|100 mA|1 A|3 A|10 A}. Default: AUTO
Example: CONF:CURR:AC 100m
8.2.1.6 CONFigure:CONTinuity
Description: Configure the instrument for continuity measurement, setting all measurement parameters and
trigger parameters to default values
Syntax: CONFigure:CONTiniuty
Parameters: (none)
Example: CONF:CONT
8.2.1.7 CONFigure:DIODe
Description: Configure the instrument for diode measurement, setting all measurement parameters and trigger
parameters to default values
Syntax: CONFigure:DIODe
Parameters: (none)
Example: CONF:DIOD
8.2.1.8 CONFigure:RESistance
Description: Configure the instrument for two-wire resistance measurement, setting all measurement parameters
and trigger parameters to default values
Syntax: CONFigure:RESistance [{<range>|AUTO|MIN|MAX|DEF}]
Parameters: <range>: {10|100|1k|10k|100k|1MA|10MA|100MA} Default: AUTO
Example: CONF:RES 10MA