EasyManuals Logo

Topcon LS-B4 User Manual

Topcon LS-B4
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
2 PREPARING FOR USE
2-9
2.3.2 Double power supply connection warning
This warning is displayed when any cable is connected to the input/output external
connector and the power is turned on while the batteries (dry batteries or
rechargeable battery pack) are installed.
When this warning is displayed, immediately turn off the power and remove the
batteries from the battery holder. Replace the empty battery holder. Turn on power
again.
Do not use the LS-B4 without the battery cover or battery holder installed.
Without it installed, water may enter the instrument and cause a malfunction.
Yellow
Red
Display
LED
LED: Blinks slowly
LS-B4E-1.book9 ページ 2001年5月18日 金曜日 午前11時59分

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon LS-B4 and is the answer not in the manual?

Topcon LS-B4 Specifications

General IconGeneral
BrandTopcon
ModelLS-B4
CategoryAccessories
LanguageEnglish

Related product manuals