39
TROUBLE SHOOTING
*1: If the error occurs frequently, contact the TOPCON support section.
*2: Refer to the instruction manual of the wireless unit (commercial product).
"Fail to execute chart software. (Error Code 607)" Turn on the switch of the power supply unit again
or restart the CV software. *1
"Fail to execute chart software. (Error Code 608)"
Connect PC-50S to the docking base with the
image cable correctly. Then, turn on the switch of
the power supply unit again. *2
"Fail to open chart software configuration screen
because the other task is in progress. Please try
again."
Display the setting screen again.
"Fail to change chart type because the other task
is in progress.
Please try again."
Change the type again.
"Fail to change chart language setting because
the other task is in progress."
Change the language again.
"Fail to open chart software configuration screen.
(Error Code 300)"
Turn on the switch of the power supply unit again
or restart the CV software. *1
"Fail to open chart software configuration screen.
(Error Code 301)"
Turn on the switch of the power supply unit again
or restart the CV software. *1
"Fail to execute chart software with selected chart
type.
Please select different chart type.
(Error Code 401)"
Turn on the switch of the power supply unit again
or restart the CV software. *1
"IDS_CHART_LANG_TYPE_ERR
(Error Code 501)"
The PC-50S software starts in the English mode.
Use the software as it is.
"Fail to save chart software configurations.
The changed parameters are canceled.
(Error Code 701)"
Turn on the switch of the power supply unit again
or restart the CV software. Then, change the set-
ting again. *1
Error message displayed on CV-5000 Remedial measure