14
PREPARATION BEFORE USE
PREPARATION BEFORE USE
INSTALLING THE INSTRUMENT
Install the instrument on the wall, using the wall-mounting metal fixture.
In addition to the wall-mounting method, you can use the Floorstand, which is an optional acces-
sory, or the stand for the commercial liquid crystal display to mount the instrument.
The interval between the screw holes in this instrument is 75mm × 75mm.
On the liquid crystal display, there is sometimes noticeably unevenness on the background due to
the disturbing light. To prevent the sunlight or disturbing light from contacting with the liquid crystal
display directly, change the installing place or direction of the instrument.
CAUTION
When installing the instrument, keep the space of 5cm or more above
and under the instrument not to block the vents on its top and bottom
surfaces. The instrument may be heated to burn someone or cause a
trouble.
CAUTION
Be careful not to damage the liquid crystal display or stain it with finger-
print, dust or others. The reliability of the measurement result may be
lowered.
CAUTION
Tighten the screws securely not to be loosened. If the screw is loos-
ened, the instrument may fall off to cause a trouble or injure someone.
NOTICE
Use the accessory screws to fix the instrument.