6 SPECIAL MODE (Menu Mode)
6.4.8 Humidity Input ON/OFF Setting
Factory setting for humidity input is "OFF". In this case, atmospheric correction factor is calculated
assuming humidity is 50%. By setting humidity input to "ON", an arbitrary value can be input for
humidity.
Humidity input:see Section 12.2 “Setting of Atmospheric Correction Value”.
Operating procedure Operation Display
1 After pressing the {MENU} key, press the {F4}
(P
↓
) key twice to get the menu on page 3.
{MENU}
{F4}
{F4}
2 Press the {F2} key.
{F2}
3 Press the {F4} key.
{F4}
4 Press the {F4} key.
{F4}
5 Press the {F2} key.
Current setting is displayed.
{F2}
6 Press the {F1} (ON) key or {F2} (OFF) key, and
press the {F4} (ENTER) key.
{F1}
or
{F2}
{F4}
MENU 3/4
F1:ILLUMINATION
F2:PARAMETERS 1
F3:PARAMETERS 2 P
↓
PARAMETERS 1 1/4
F1:MINIMUM READING
F2:AUTO POWER OFF
F3:TILT P
↓
PARAMETERS 1 2/4
F1:ERROR CORRECTION
F2:RS-232C
F3:SELECT COM P
↓
PARAMETERS 1 3/4
F1:BLUETOOTH
F2:HUMID.INPUT
F3:NP-TRK. MODE P
↓
HUMID.INPUT [ON ]
F1:ON
F2:OFF
ENTER