P2.4
P2.5
0 Select filter level against sudden impacts.
P2.6
ADC Internal
Filter
0 100 sps.
1 50 sps.
2 25 sps.
3 12.5 sps.
4 6.25 sps.
P2.7
Standstill
Detection
Interval
1 1e
2 2e
3 3e
4 4e
5 5e
1 If the difference between consecutive weight
values is smaller than this value, standstill
signal will light after the time period selected
at P2.7 is elapsed.
P2.8
Standstill
Waiting
Time
0 0 sec.
1 0,5 sec.
2 1 sec.
3 1,5 sec.
4 2 sec.
2 Select time period
will be elapsed
before
standstill signal lights
(
if the consecutive
weight values are smaller than P2.6).
P2.9
Tar e
Request
0 Tare Enable
1 Tare Disable
0 Tare button can be enabled o r disabled.
KM.LL2E.002
Load Cell
Gain Rate
1 x1
2 x2
4 x4
8 x8
16 x16
32 x32
64 x64
64
Select gain rate of mV sig nal coming from
load cells.
Select 64 for 1 mV/V load cells,
Select 64 or 32 for 2mV/V load cells,
Select 64,
32 or
16 for others.
1, 2, 4 and 8 are dedicated for special
applications.
Digital Filter 0 Disable
1 Last 2
2 Last 4
3 Last 8
4 Last 16
5 Last 32
5 Select filter level. Moving average of last 2
32 weight vaules will be calculated.