9. ERROR MESSAGES
Error 1 ADC Conversion Error
Restart the device and call our firm if error message
appears again
Error 2 Signal Failure Signals are oppositely connected. Correctit..
Error 3 Weak Signal
Check calibration parameters (especially P3.8) and
steel construction of the scale.
Error 4 Division Number Error
Check maximum capacity and verification scale
intervals.
Error 5 Initialization Error
Restart device. Be sure that load is still while the
device is opening.
Error 6 Calibration Weight Error
Calibration Weight must be greater than or equal to
5% of maximum capacity.
Error 7 Parameter Error Check all parameters.
Error 8
Incorrect Calibration
Password
Check your calibration password and then try again.
Error 10
Calibration jumper does
not exist
Attach calibration jumper. C alibration jumper must
be attached in order to enter some of programming
and calibration parameters.
Error 11 Date Time Error Check date and time parameters.