23
P4.1
Self
Calibration
-------
P4.2
Gravity
Acceleration
------- Gravity acceleration parameters make it
possible to compensate the
weight difference
between the place in which the instrument is
calibrated (GR-CAL) and the p
lace in which
the instrument will be used (GR-USE)
, due to
different gravity acceleration.
Gravity acceleration values must be between
9,77000 and 9,84000 Enter these two
parameters by using keypad (starting from
left) and press enter button.
A
fter that
the
,
weight
corrected.
calibration is automatically
Gravity values at some places are as follows:
2
Amsterdam 9.813 m/s
Athens
Auckland NZ
Bangkok
Birmingham
Brussels
Buenos Aires
Calcutta
Chicago
Copenhagen
Djakarta
Frankfurt
Glasgow
Havana
Helsinki
Kuwait
Lisbon
London (Greenwich)
Los Angeles
Madrid
2
9.800 m/s
2
9.799 m/s
2
9.783 m/s
2
9.813 m/s
2
9.811 m/s
2
9.797 m/s
2
9.788 m/s
2
9.803 m/s
2
9.815 m/s
2
9.781 m/s
2
9.810 m/s
2
9.816 m/s
2
9.788 m/s
2
9.819 m/s
2
9.793 m/s
2
9.801 m/s
2
9.812 m/s
2
9.796 m/s
2
9.800 m/s
This method is used to calibrate the device
without calibrated weigh ts. You must enter
number of load cells (Lc No),
capacity of load
cells (Lc Cap), mi llivolt signal (Lc Sens) and
to let the device sense the dead load of the
scale (No Load) . Finally press Enter
button,
device will show “BUSY” message.
Note: self calibration can be used only for not
legal for trade applications
KM.LL2E.002