2.2 Safety features incorporated in the device
The pressure vessel chamber door has the following features protecting
personnel from hazards:
l Two door switches indicate that the door is closed. Without this indication
steam is not introduced into the chamber.
l An electrical door locking pin that blocks door opening during operation.
The following safety devices are installed in the autoclave to optimize its safe
operation:
l Safety thermostat, to prevent over- heating of the chamber heating
elements.
l Safety cut-off switch to prevent over-heating of the pipe heating element.
l Safety pressure valve to prevent over-pressurizing of the chamber.
Warning!
Mind the Power socket. Keep it and its vicinity dry.
Danger of electrocution.