8.3 Starting a Cycle
It is recommended to perform B&D test cycle at the beginning of each working
day.
1. Before each cycle, check visually that the gasket is intact, not loose and
clean.
2. Load the autoclave properly (see chapter 6).
3. Choose the appropriate sterilization program.
4.
Each sterilization program screen contains a Virus Protect option.
Notes:
l The program can only be selected when the door is open.
l The Virus Protect is not for use for Textile and Porous programs
5. The selected programs are shown below:
Class B
Class S
See section 8.5 for available sterilization and test programs.
Page 47