Message Description
Please fill water tank to full
for start
For prion cycle in manual water filling
configuration, when the water tank is not
full the “Start” button will not appear,
instead this error message will appear in
the Active alarms screen.
Prion cycle, in manual water filling and
the water tank is not full
Unrecognized printer
If printer is not recognized an error will
appear in the UI.
Low Quality Water
1. Read the water level from analog
sensor.
2. Translate the digital value to micro
siemens according to the attached
table.
3. Compare the value system
parameter (water quality
reference, default 10).
4. if value is less than 10 show error -
Low Quality Water
Set Atmosphere Pressure
is Active
Present when Reset Atmospheric
pressure is requested from Technician
screen
“Not enough water
detected”
If in the clean water tank, the Float switch
does not change from 1 to 0 during water
pump time On
Pipe water heater
temperature time error
If the Pipe Water Heater works for more
than 15 min.
Jacket water heater
temperature time error
If the Jacket Water Heater works for more
than 15 min.
“Analog Input Error - Jacket
Temperature”