KANGAROO JOEY
8
Enteral Feeding Pump
(Note: display status line should read SET LOADED).
4. Suspend the pump set bag so that the top of the starting volume of formula is 6 inches above the pump (Figure
4B).
5. To automatically prime the pump, press
4PRIME PUMP and4AUTO PRIME. For Feed & Flush sets, auto priming will prime
both lines.
6. To directly control pump priming, press
4PRIME PUMP and then press and hold the HOLD TO PRIME button(s)4. If a Feed
& Flush Set is used, remember to first press4HOLD TO PRIME FLUSH until the flushing fluid has reached the valve, and then
press4HOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the distal end connector.
7. Use buttons
4on left side of LCD to set the feeding parameters. Set flushing parameters if a Feed & Flush Set is loaded. The
continuous or intermittent feed mode selection is made under the MORE options menu.
8. Press
4RUN when ready to start. The screen will display RUNNING.
9. To stop, press4HOLD or press and hold the POWER button.
182541
6 in
(15.2 cm)
6 in
(15.2 cm)
Figure 4B. Correct Fluid Level.
Figure 4A. Installation of Pump Set. (See Figure 6 for detailed view.)
WARNING: DO NOT OVERSTRETCH TUBE.
1
3
2
Black Ring
Retainer
(MISTIC)
Finger
Tab
Valve
Rotor
General Startup
Placement/Mounting
There are two recommended placement methods for the KANGAROO JOEY Enteral Feeding Pump.
• Attached to a vertical IV pole via the pole clamp peripheral device, included with the pump.
• Placed on any stable surface.
A/C Power Operation
Plug the KANGAROO JOEY Enteral Feeding Pump into an A/C power outlet for normal operation. When the pump is not plugged in,
or if A/C power is interrupted, a built-in rechargeable battery will run the pump.
Battery Power Operation
If A/C power is not available to the pump, or if A/C power is interrupted, the pump will automatically operate on backup battery
power. The built-in battery recharges any time the pump is plugged into an outlet.